Standard continuous monitoring in CCU uses 5 electrodes providing leads I, II, III, aVR, aVL, aVF, and one precordial lead.
Right arm / below clavicle R
Mnemonic
Left arm / below clavicle L
Right leg / lower abdomen R
V1: 4th ICS right sternal border
V5/V6: lateral chest
Left leg / lower abdomen L
| Lead | Best For | Clinical Use |
|---|---|---|
| Lead II | Best P-wave visibility; inferior MI detection | Default continuous monitoring lead |
| V1 / MCL1 | RBBB vs LBBB differentiation; PVC vs aberrant conduction | Selected when bundle branch morphology matters |
| V5/V6 | Lateral ischaemia, ST changes | Post-PCI monitoring, lateral STEMI surveillance |
Continuous (Bedside/Telemetry)
Ongoing real-time rhythm surveillance. Detects rate changes, arrhythmias, ST trends. Uses 3–5 leads simultaneously. Does NOT replace a 12-lead ECG.
12-Lead ECG
Snapshot recording of cardiac electrical activity from 12 perspectives. Required for ischaemia diagnosis, bundle branch block analysis, QT interval measurement, and any new symptom (chest pain, syncope, dyspnoea).
| Hardwire | Telemetry | |
|---|---|---|
| Location | Bedside only | Ward / mobile |
| Signal | Wired cable | Radio frequency |
| Range | No movement | Typically 30–100 m |
| Indications | ICU/CCU, critically ill, haemodynamically unstable | Step-down, post-cath, stable arrhythmia, post-arrest observation |
| Signal loss | Rare (cable fault) | Interference, body position, distance |
| Artifact Type | Likely Cause | Solution |
|---|---|---|
| Baseline wander | Patient movement, respiration, poor electrode adhesion | Reposition electrodes away from muscle; ensure skin clean and dry |
| 60 Hz / AC interference (thick fuzzy line) | Electrical equipment nearby, electrode disconnection | Check all connections; remove nearby electrical devices; replace electrodes |
| No signal / flat line | Lead disconnection, dried electrode gel, lead fracture | Verify connections; replace electrode; check cable integrity |
| Muscle tremor artefact | Shivering, Parkinson's, patient anxiety | Warm patient; reposition electrodes to bony prominences |
| False alarms | Poor contact mimicking VF/asystole; motion mimicking VT | Replace electrodes; reassess skin prep; adjust alarm thresholds per policy |
Priority Alarms (Respond Immediately)
- VF / pulseless VT
- Asystole / extreme bradycardia (<30 bpm)
- 3rd degree heart block
- Pacemaker failure to capture in pacemaker-dependent patient
- QTc >500 ms (new)
- ST change >1 mm from baseline
Alarm Reduction Strategies (GCC Policy)
- Change electrodes every 24–48 hours (or per unit protocol)
- Clean and dry skin before electrode application
- Set individualised thresholds — not default factory settings
- Document reason for alarm threshold changes
- Conduct multi-disciplinary alarm safety huddles
- Telemetry watcher (central monitoring) roles in larger GCC hospitals
-
RateRegular rhythm: 300 ÷ number of large squares between R-R (e.g., 4 squares = 75 bpm)
Irregular rhythm: Count R waves in 6-second strip × 10
Normal: 60–100 bpm | Tachy: >100 | Brady: <60 -
RegularityMeasure consecutive R-R intervals with calipers or paper:
Regular — all R-R equal | Regularly irregular — pattern (e.g., Wenckebach) | Irregularly irregular — no pattern (AF) -
P WavesPresent? Upright in II? One P before every QRS? Same morphology?
Absent P waves → AF, junctional. Sawtooth → atrial flutter. Retrograde P → junctional. -
PR IntervalNormal: 0.12–0.20 s (3–5 small squares at 25 mm/s)
>0.20s = 1st degree block | Progressive lengthening = Wenckebach | Fixed prolonged with dropped QRS = Mobitz II -
QRS WidthNormal: <0.12 s (<3 small squares)
Wide QRS ≥0.12s → BBB, WPW, paced rhythm, ventricular origin (VT, PVC)
| Measurement | Normal Range | Small Squares (25 mm/s paper) |
|---|---|---|
| PR interval | 0.12 – 0.20 s | 3 – 5 squares |
| QRS duration | < 0.12 s | < 3 squares |
| QT interval | < 0.44 s (male) / <0.46 s (female) | < 11 squares |
| QTc >500 ms | Critical — risk of Torsades de Pointes; review QT-prolonging drugs | |
| P wave duration | < 0.12 s | < 3 squares |
Insertion Sites (Preferred Order)
- Radial artery — first choice; perform Allen's test first
- Femoral artery — large vessel; higher infection risk
- Brachial artery — avoid if possible; end-artery in some
- Dorsalis pedis, ulnar — alternatives
Allen's Test (Pre-Radial A-Line)
- Occlude both radial and ulnar arteries (patient makes fist)
- Release ulnar only — hand should flush pink within 5–7 seconds
- >15 seconds = ABNORMAL — collateral circulation inadequate; choose alternative site
Arterial Waveform Components
- Systolic upstroke — rapid ejection; slope reflects contractility
- Systolic peak — SBP reading
- Dicrotic notch — aortic valve closure; landmark on downstroke
- Diastolic runoff — DBP at lowest point
Zeroing Procedure
- Position TransducerLevel transducer stopcock at the phlebostatic axis — 4th intercostal space, mid-axillary line (right atrial level). Use levelling device or laser.
- Open to AirTurn stopcock off to patient; open stopcock to atmosphere. Ensure no movement of patient or bed.
- Zero on MonitorPress "Zero" on bedside monitor. Confirm waveform reads 0 mmHg. Close stopcock to air; reopen to patient.
- DocumentDocument time of zeroing. Re-zero after patient position changes, after circuit manipulation, or if readings are inconsistent.
Waveform Damping
| Type | Cause | Effect on Readings | Action |
|---|---|---|---|
| Overdamped | Air bubbles, clot, kink, loose connection | Falsely LOW systolic; falsely HIGH diastolic; blunted waveform | Flush system; check for clot; remove air; check connections |
| Underdamped | Long tubing, stiff tubing, excessive tapping | Falsely HIGH systolic; falsely LOW diastolic; narrow spike appearance | Shorten tubing; use compliant tubing; minimise movement |
Complications Monitoring
Check connections regularly; never leave stopcock open; assess insertion site; ensure Luer-lock connections.
Hourly distal pulse check; assess colour, temperature, sensation of distal limb; document Allen's test result.
Sterile dressing changes per policy; assess site for redness/pus; change circuit per hospital protocol (typically 72–96 h).
Normal Range
2 – 8 mmHg
CVP reflects right atrial pressure and is a crude estimate of preload. Trend is more valuable than single reading.
| CVP Trend | Interpretation |
|---|---|
| <2 mmHg | Hypovolaemia; fluid challenge may be indicated |
| 2–8 mmHg | Normal; correlate with clinical picture |
| >12 mmHg | Right heart failure, cardiac tamponade, fluid overload, PEEP effect |
Mean Arterial Pressure = (SBP + 2 × DBP) ÷ 3
Target MAP ≥65 mmHg in sepsis / cardiogenic shock
Pulse Pressure Variation (PPV)
PPV = (PPmax − PPmin) / PPmean × 100
PPV >13% in mechanically ventilated patients suggests fluid responsiveness.
| Method | Principle | Nursing Role |
|---|---|---|
| PA Catheter (Swan-Ganz) Thermodilution | Cold saline bolus → temperature-time curve → CO calculation | Administer cold saline consistently; document CO, CI, PCWP; watch for arrhythmias on insertion; monitor balloon deflation |
| PiCCO Transpulmonary thermodilution | Calibration via central and arterial line; continuous CO via pulse contour | Recalibrate per protocol (position change, haemodynamic instability); ensure arterial line positioned correctly |
| USCOM Ultrasound cardiac output | Doppler measurement of aortic or pulmonary flow velocity | Non-invasive; nurse-led in some GCC centres; requires training; document CO and VTI values |
Pacemaker Settings
| Setting | Function | Typical Value |
|---|---|---|
| Rate | Pacing rate (bpm) | 60–80 bpm |
| mA (Output) | Electrical current delivered; ensures capture | 2–3× capture threshold |
| Sensitivity (mV) | Threshold for sensing native R waves | Start 2–5 mV; adjust as needed |
| Mode | Demand (VVI) or asynchronous (VOO) | Demand (VVI) default |
Paced Rhythm Recognition
- Pacemaker spike (vertical line) precedes each paced complex
- Capture confirmed: spike followed by QRS (ventricle) or P wave (atrium)
- Paced QRS is typically wide and bizarre (LBBB morphology for RV pacing)
- In demand mode: native beats should inhibit pacing — no spike on native QRS
Troubleshooting Pacemaker Problems
| Problem | ECG Finding | Cause | Action |
|---|---|---|---|
| Failure to Capture | Spike present but no QRS follows | Lead dislodgement, output too low, exit block, battery failure | Increase mA; check lead connections; reposition patient; notify doctor; check battery |
| Failure to Sense | Spikes on native beats (competition) | Sensitivity set too high (mV too high = less sensitive); lead fracture | Decrease mV setting (more sensitive); check lead; switch to asynchronous if emergency |
| Failure to Pace | No spikes when expected | Battery dead, generator fault, lead fracture, inhibition by interference | Replace battery; check connections; eliminate electromagnetic sources; change generator |
| Oversensing | Pacing inhibited by noise (T wave, muscle) | Sensitivity too low (mV too low = too sensitive); electromagnetic interference | Increase mV; remove interference sources; consider asynchronous mode |
PPM Checks (Nursing)
- Verify programmed rate on ECG (usually 60 ppm lower rate)
- Check pacemaker ID card on admission
- Document pacemaker manufacturer, model, implant date, programmed mode (e.g., DDD, VVI)
- MRI conditional vs non-conditional: verify before any MRI
- Avoid placing BP cuff on arm ipsilateral to subclavian lead
- Annual/biannual device checks (remote monitoring in many GCC hospitals)
ICD — Appropriate vs Inappropriate Shocks
| Type | Trigger | Action |
|---|---|---|
| Appropriate | VT/VF detected correctly | Post-shock 12-lead ECG; document; notify cardiologist; assess for haemodynamic stability |
| Inappropriate | AF/SVT/noise sensed as VT/VF | Notify cardiologist urgently; device reprogramming may be needed; do NOT deactivate magnet without doctor order |
| Cardioversion (Synchronised) | Defibrillation (Unsynchronised) | |
|---|---|---|
| Indication | AF, atrial flutter, SVT, stable VT with pulse | Pulseless VT, VF |
| Synchronisation | YES — shock delivered on R wave to avoid T wave (R-on-T) | NO — immediate shock regardless of cardiac cycle |
| Sedation | Required (conscious patient) — midazolam, propofol, or ketamine | Not required (patient unconscious/pulseless) |
| Energy (Biphasic) | AF: 100–200 J | Flutter: 50–100 J | SVT/VT: 50–150 J | First shock: 150–200 J; subsequent: 200–360 J (device-dependent) |
| Pre-procedure | Anticoagulation ≥3 weeks if AF >48 h (or TOE to exclude thrombus); consent; fasting | No preparation needed — emergency |
AED — Guiding a Bystander
- Turn on AED — follow voice prompts
- Attach pads: right of sternum below clavicle; left lateral chest below armpit
- Paediatric (<8 years / <25 kg): use paediatric pads; if unavailable, place one pad anteriorly, one on back (anterior-posterior)
- Analyse: "Do not touch patient"
- If shock advised: "Stand clear" → shock button → immediately resume CPR for 2 minutes
- If no shock advised: immediately resume CPR
Hypokalaemia (<3.5 mmol/L): AF, VT, U wave on ECG, prolonged QU interval, increased digoxin toxicity risk
Hyperkalaemia (>5.5 mmol/L): peaked T waves, wide QRS, sine wave pattern, VF, asystole
Target in CCU: K⁺ ≥4.0 mmol/L post-MI
Hypomagnesaemia (<0.75 mmol/L): Refractory arrhythmias (resistant to treatment until Mg corrected); Torsades de Pointes; often co-exists with hypokalaemia
Treatment: MgSO₄ 2 g IV for Torsades; replace simultaneously with K⁺
Hypocalcaemia: Prolonged QT interval → Torsades risk; prolonged ST segment
Hypercalcaemia: Short QT interval, shortened ST; bradycardia; 1st/2nd degree block; can sensitise to digoxin toxicity
Action: Check ionised Ca²⁺; replace cautiously
- AmiodaroneBradycardia (sinus node suppression); QT prolongation; thyroid dysfunction can exacerbate arrhythmias; monitor TFTs, LFTs, CXR
- Digoxin ToxicityClassic: PAT with 2:1 block; bidirectional VT; AV nodal blocks; early signs: nausea, yellow/green visual halos; therapeutic range 0.5–2 ng/mL; K⁺ depletion potentiates toxicity
- SotalolQT prolongation → Torsades; more pro-arrhythmic than other beta-blockers; dose-dependent QTc effect
- Macrolide antibiotics (erythromycin, azithromycin)QT prolongation via hERG K⁺ channel blockade; significant concern in elderly and with polypharmacy
- Antipsychotics (haloperidol, quetiapine)QT prolongation; risk increases with IV haloperidol; always check QTc before administration in CCU
- Fluconazole / VoriconazoleQT prolongation; inhibit CYP3A4 increasing levels of co-administered QT-prolonging drugs
- Cisplatin / CyclophosphamideCardiotoxicity; arrhythmias; electrolyte depletion (Mg, K) exacerbating risk
- Tricyclic Antidepressants (TCA)Overdose: wide complex tachycardia, prolonged QRS and QT; treat QRS widening with sodium bicarbonate IV
Rhythm Strip Labelling Requirements
- Date and time of print
- Patient full name, MRN, date of birth
- Ward / bed number
- Lead(s) displayed
- Nurse interpretation (rhythm, rate)
- Action taken
- Nurse signature and designation
When to Print a Rhythm Strip
- Each nursing shift (per policy — typically 8–12 hourly)
- Any new arrhythmia or rhythm change
- Haemodynamic change (BP drop, altered consciousness)
- Alarm triggered by rhythm change
- After cardioversion / defibrillation
- Pacemaker malfunction or threshold testing
DHA/DOH Telemetry Competency Requirements
• Formal ECG/telemetry certification (hospital-based or external provider)
• Annual competency renewal
• ACLS/BLS currency
• Demonstrated rhythm recognition and response
• Documentation of in-service training hours
Telemetry Watcher Role (Central Monitoring)
- Dedicated monitoring of multiple patients from central station
- Immediate communication to bedside nurse for any critical alarm
- Document all rhythm changes with time stamp
- Cannot replace bedside nurse clinical assessment
Q1. A patient's ECG shows an irregularly irregular rhythm, no identifiable P waves, and narrow QRS complexes at a rate of 110 bpm. What is the MOST likely rhythm?
Q2. Your pacemaker-dependent patient has a pacemaker spike on the ECG but no QRS complex following it. What is this called and what is your FIRST action?
Q3. A patient's potassium is 2.9 mmol/L. Which arrhythmia is this patient at INCREASED risk of developing?
Q4. Before inserting a radial arterial line, the nurse performs an Allen's test. The patient's hand does not flush pink until 18 seconds after releasing the ulnar artery. What should the nurse do?
Q5. A conscious patient with known AF (onset 6 hours ago) develops chest pain and a BP of 75/40 mmHg. What is the MOST appropriate immediate intervention?
Q6. An ECG shows progressively lengthening PR intervals followed by a dropped QRS beat, then the pattern repeats. This is:
Q7. You note that an arterial line waveform appears blunted with a falsely low systolic reading. This is most consistent with:
Q8. A patient with a QTc of 520 ms is prescribed IV haloperidol for agitation. What should the nurse do?
Q9. During a defibrillation attempt, a team member says "I'll stay connected to the IV line — it's just a drip." What is your response?
Q10. A patient's monitor shows a wide-complex tachycardia at 180 bpm. The patient is conscious, BP 90/60 mmHg, and diaphoretic. What is the correct initial classification of this rhythm and your priority action?